Class II
Medical Device Recall: Abbott Navitor Transcatheter Aortic Heart Valve, 29 mm, NVTR-29, Sterile
Abbott Medical · October 17, 2024
Reason for Recall
A manufacturing error resulting in acceptance of valves with a leaflet deflection value that was outside of established specifications. This could potentially affect long-term durability of the impacted valve and/or cause potential valve failure.
Distribution
New York, Australia, Belgium, Germany, Iran, Israel, Japan, Kazakhstan, Oman, Poland, Romania, Slovakia, Spain, Turkey, United Kingdom.
Quantity Affected
1 OUS