Medical Device Recall: Merlin" Patient Care System (PCS) software Merlin" PCS 3650 programmer Model 3330 software
Abbott Medical · May 12, 2026
Reason for Recall
Software malfunction may occur during pacing capture threshold test of pacemaker resulting in programmer not successfully communicating the command to terminate the test.
Distribution
US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY. OUS: Not provided yet
States Affected
NATIONWIDE
Quantity Affected
11182 programs
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2405-2026
Status: ongoing