Class II

Medical Device Recall: The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates using the Merlin PCS Model 3330 software and is designed

Abbott Medical · November 21, 2024

Reason for Recall

Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the finalization step to fail. If this occurs, the Programmer displays a loss of telemetry detected window, and there is no pacing.

Product Description

The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates using the Merlin PCS Model 3330 software and is designed to interrogate, program, display data from, and test Abbott Medical implantable devices and leads.

Distribution

Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN TX, UT, VA, WA, WI, and WV. The countries of Austria, Belgium, Canada, Denmark, Finland, France, Germany, Hong Kong, Israel, Italy, Kuwait, Netherlands, Norway, Poland, Portugal, Saudi Arabia, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom.

States Affected

NATIONWIDE

Quantity Affected

7295 programmers

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0753-2025

Status: ongoing