Class II

Medical Device Recall: Abiomed Axillary Insertion Introducer -Introducer(s) are intended for introduction of the Impella Catheter into the body. Part Number: 0052-3006 Individual, and Part of Pump Set (005062, 0550-0008,

Abiomed, Inc. · November 7, 2023

Reason for Recall

Hole(s) to the outer pouch of the Introducer Kits (both trayed and non-trayed) may compromise the sterile barrier of the Introducer Kits. Non-sterile product exposes patients to the possibility of the introduction of micro-organisms into the vasculature and/or access site, leading to an infectious process, bacteremia or sepsis

Product Description

Abiomed Axillary Insertion Introducer -Introducer(s) are intended for introduction of the Impella Catheter into the body. Part Number: 0052-3006 Individual, and Part of Pump Set (005062, 0550-0008, 1000100, 0052-0011, 1000220)

Distribution

Nationwide Foreign: Country AT AU BE BR CA CA CH CZ DE DK ES FR FR HK IN IT MX MY NL NO PA SG TW UK

States Affected

NATIONWIDE

Quantity Affected

717 single units; 1906 within sets

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0646-2024

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.