Medical Device Recall: GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317. A required kit for the GenetiSure Dx Postnatal assay (K1201A).
Agilent Technologies, Inc. · March 20, 2025
Reason for Recall
DNA Labeling Kit for use in sample labeling to be hybridized to oligonucleotide microarrays has a component, Human Reference DNA Female aliquot, affected by a mosaic aberration that may result in Chromosome 12 baseline shift, which may lead to false positive/negative results, resulting in diagnosis delay, and may lead to delayed treatment of a treatable genetic disorder.
Distribution
US distribution to the state of: MA and OUS (International) distribution to countries of: Spain, Italy, Japan, Colombia, Poland
States Affected
MA
Quantity Affected
70
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1544-2025
Status: ongoing