Class II

Medical Device Recall: Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware

Agiliti Health - Ellis · January 21, 2026

Reason for Recall

Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.

Product Description

Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

4286

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1472-2026

Status: ongoing