Class II

Medical Device Recall: Brand Name: Alcon Product Name: Monarch¿ III D Delivery Cartridge Model/Catalog Number: 8065977763 Product Description: Monarch¿ III D delivery cartridge Lens Guide, Intraocular Component: N/A

Alcon Research LLC · November 13, 2024

Reason for Recall

Increase in complaints associated with a molding defect within the inner lumen of delivery cartridges which could result in a scratch on the intraocular lens (IOL) upon delivery.

Distribution

US: AK AL AR AZ CA CO CT DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA RI SC SD TN TX UT VA VT WA WI WV WY OUS: Russia, Egypt

States Affected

AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY

Quantity Affected

41718 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0835-2025

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.