Medical Device Recall: Brand Name: Alcon Product Name: Monarch¿ III D Delivery Cartridge Model/Catalog Number: 8065977763 Product Description: Monarch¿ III D delivery cartridge Lens Guide, Intraocular Component: N/A
Alcon Research LLC · November 13, 2024
Reason for Recall
Increase in complaints associated with a molding defect within the inner lumen of delivery cartridges which could result in a scratch on the intraocular lens (IOL) upon delivery.
Distribution
US: AK AL AR AZ CA CO CT DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA RI SC SD TN TX UT VA VT WA WI WV WY OUS: Russia, Egypt
States Affected
AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY
Quantity Affected
41718 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0835-2025
Status: ongoing