Medical Device Recall: Surgical Image Guidance Functionality, software used with NGENUITY 3D Visualization System, Part Number: 8065000399
Alcon Research LLC · May 23, 2024
Reason for Recall
After surgery initiation, while using surgical image guidance, if significant eye rotation/movement occurs, a software anomaly may cause re-establishment of incorrect new registration angle, range indicator will display incorrectly, which could result in incorrect placement of toric IOL axis, which could cause astigmatic error under/over correction resulting in decreased uncorrected visual acuity.
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, CA, CO, CT, FL, GA, HI, IA, IL, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NJ, NV, NY, OH, OK, PA, SC, TN, TX, WA, WI, PR and the countries of Canada, Japan, France.
States Affected
NATIONWIDE
Quantity Affected
230
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2412-2024
Status: ongoing