Class II

Medical Device Recall: Custom procedural convenience kits and trays, cardiac, labeled as: a) CENTRAL LINE INSERTION PACK, kit number LLCE52; b) Adult Central Line Insertion Supply Kit, kit number UICL44N; c) Pediatri

American Contract Systems, Inc. · November 29, 2023

Reason for Recall

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Product Description

Custom procedural convenience kits and trays, cardiac, labeled as: a) CENTRAL LINE INSERTION PACK, kit number LLCE52; b) Adult Central Line Insertion Supply Kit, kit number UICL44N; c) Pediatric Central Line Insertion, kit number UIPL77K

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

290 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0819-2024

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.