Class II

Medical Device Recall: Custom procedural convenience kits and trays, Obstetrical & Gynecological use, labeled as: a) LavH, kit number BBLV21C; b) LAP HYST 247540, kit number BFHY75U

American Contract Systems, Inc. · November 29, 2023

Reason for Recall

During an internal investigation, ACS identified that several components were "piggybacked" to ACS trays and subjected to the ACS sterilization process. There is no supporting documentation stating the "piggybacked" components were assessed to determine if they were suitable for exposure to elevated temperatures for an extended period of time and further exposure to sterilization. As a result, the piggybacked components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) may have had higher than specified EO residuals.

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

144 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0831-2024

Status: ongoing