Medical Device Recall: Medical convenience kits Item Number/Description HSAN63N ANGIO PACK IHAN02W ANGIOGRAPHY PACK MHCA98AK CORONARY ANGIO PACK- 205991 RCAN45C ANGIO CUSTOM RCH UDCO10F CVL CORONARY PACK 219224

American Contract Systems Inc. · September 4, 2025

Reason for Recall

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Distribution

US Nationwide distribution in the states of SD, IA, MN, WA, IL.

States Affected

NATIONWIDE

Quantity Affected

32,433 total

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0180-2026