Class I

Medical Device Recall: Arrow FiberOptix Intra-Aortic Balloon Catheter Kit: a) REF IAB-05830-LWS; b) REF IAB-05840-LWS; c) REF IAB-05850-LWS

ARROW INTERNATIONAL Inc. · April 29, 2024

Reason for Recall

Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.

Distribution

Worldwide distribution. US nationwide including Puerto Rico, AU, BA, BE, BG, BN, CA, CL, CN, CO, CY, CZ, DE, EC, ES, FR, GB, GP, GR, HU, HK, IE, IL, IN, IQ, IT, JO, JP, KW, KZ, LB, LU, MA, MN, MK, MX, MY, NL, OM, PA, PF, PH, PL, PT, QA, RE, RO, RS, SA, SE, SG, SI, SK, TH, TR, TW, UZ, ZA, and UAE.

States Affected

NATIONWIDE

Quantity Affected

7939 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1878-2024

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.