Class II

Medical Device Recall: Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 89 cm (35"), REF US530, to prevent or treat hypothermia

Augustine Temperature Management, LLC · December 16, 2024

Reason for Recall

There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.

Distribution

US Nationwide distribution in the states of TX, AL, NV; AU, NZ.

States Affected

NATIONWIDE

Quantity Affected

3 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1199-2025

Status: ongoing