Medical Device Recall: MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006 , MON-000006-RFB, MON-000008 with software
Auris Health, Inc · January 21, 2026
Reason for Recall
Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.
Distribution
Worldwide - US Nationwide distribution in the states of MN, FL, NJ, NY, VA, WI, AZ, AL, MS, LA, MI, MA, MT, PA, WV, GA, DE, TX, CA, OH, IN, SC, CT, ND, MO, DC, IL, NE, ID, WA, NC, TN, OR, CO, SD, NM, VT, MD and the countries of Hong Kong, China, Canada.
States Affected
NATIONWIDE
Quantity Affected
173
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1383-2026
Status: ongoing