Class II

Medical Device Recall: Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Software Version: 8713031U Product Description: Perfusor Space Syringe Pump, Infusion

B Braun Medical Inc · June 14, 2024

Reason for Recall

Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.

Product Description

Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Software Version: 8713031U Product Description: Perfusor Space Syringe Pump, Infusion Pump, 1 each Component: N/A

Distribution

Domestic distribution nationwide. International distribution to Canada.

States Affected

NATIONWIDE

Quantity Affected

9943

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2434-2024

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.

B Braun Medical Inc Medical Device Recall: Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Software Version: 8713031U Product Description: Perfusor Space Syringe Pump, Infusion | SafeCheck