Medical Device Recall: Brand Name: Perfusor Space Product Name: Perfusor PCA Syringe Pump Model/Catalog Number: 8713080U Product Description: Perfusor¿ Space Syringe Pump, Infusion Pump, 1 each Component: N/A
B Braun Medical Inc · June 14, 2024
Reason for Recall
Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
Distribution
Domestic distribution nationwide. International distribution to Canada.
States Affected
NATIONWIDE
Quantity Affected
1663
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2435-2024
Status: ongoing