Class II

Medical Device Recall: SL BLOODLINE FOR FMC 2008 SERIES-Tubing to be used to transfer blood between a patient and a haemodialyser during haemodialysis. Model/Catalog Number: SL-2000M2095

B Braun Medical Inc · August 15, 2024

Reason for Recall

Incomplete insertion of the patient connector of the Streamline Airless System Hemodialysis Bloodline to the Locksite may result in the potential for occlusion during prograde rinseback procedures. If the patient connector is not fully inserted into the Locksite, the internal septum may not open fully enough to prevent restrictions to flow. may cause delays in therapy while clinicians are forced to switch to retrograde flow method

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

10658351 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-3173-2024

Status: ongoing