Class II

Medical Device Recall: Bard Mission Disposable Core Biopsy Instrument Kit (REF/Name): 1410MSK/ Mission Kit 14G x 10CM; 1416MSK/ Mission Kit 14G x 16CM; 1610MSK/ Mission Kit 16G x 10CM; 1616MSK/ Mission Kit 16G x 16CM; 1810M

Bard Peripheral Vascular Inc · November 8, 2023

Reason for Recall

The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy needle may not fit properly into the coaxial cannula, preventing access to the target tissue, which could lead to tissue injury, longer and repeat procedures.

Product Description

Bard Mission Disposable Core Biopsy Instrument Kit (REF/Name): 1410MSK/ Mission Kit 14G x 10CM; 1416MSK/ Mission Kit 14G x 16CM; 1610MSK/ Mission Kit 16G x 10CM; 1616MSK/ Mission Kit 16G x 16CM; 1810MSK/ Mission Kit 18G x 10CM; 1816MSK/ Mission Kit 18G x 16CM; 1820MSK/ Mission Kit 18G x 20CM; 1825MSK/ Mission Kit 18G x 25CM; 2010MSK/ Mission Kit 20G x 10CM; 2016MSK/ Mission Kit 20G x 16CM; 2020MSK/ Mission Kit 20G x 20CM

Distribution

Worldwide - US Nationwide distribution including in the states of PR, IL, MA, GA, KY, CA, TX, PA, IA, MN, NH, MD, FL, ID, IN, AL, VA, CO, WI, NY, NV, OK, OH, WA, MI, MS, AZ, HI, SD, NJ, NE, MT, NC, DC, CT, WY, UT, TN, LA, NM, SC, VT, MO, ME, AR, OR, KS, RI, AK, WV, DE, ND, PR, GU and the countries of CA, BE, AU, KR, SG, CN, TH, HK, VI, VN, AR, JP, CL, IN, TW, CO, MX, DO, BR, BD, MM, PK, NP, PA.

States Affected

NATIONWIDE

Quantity Affected

1,243,937