Medical Device Recall: Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual Lumen Hemodialysis Catheters. Please refer to the applicable catheter Instruction
Bard Peripheral Vascular Inc · July 10, 2024
Reason for Recall
Due to guidewires being manufactured incorrectly with the guidewires being mis-assembled within the hoop such that the proximal (stiff) end of the guidewire is located at the distal end of the hoop. Injury could occur if the user attempts to insert the proximal end into the patient.
Product Description
Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual Lumen Hemodialysis Catheters. Please refer to the applicable catheter Instructions for Use for directions on the use of these accessories.
Distribution
Worldwide - U.S. Nationwide distribution in the states of OH, TX, and WV. The countries of Belgium, Canada, France, Italy, and United Kingdom.
States Affected
NATIONWIDE
Quantity Affected
625 units
Source: FDA (openFDA Device Enforcement)