Class II

Medical Device Recall: Safe-T-Centesis 6 Fr Catheter Drainage Tray, REF: PIG1260T,

Bard Peripheral Vascular Inc · November 6, 2025

Reason for Recall

Acute drainage catheters, indicated for use in removing fluid during acute drainage/aspiration procedure, manufactured with safety indicators that erroneously show catheter in sharp position when catheter is in blunt position, which may lead users to believe device is not yet in pleural or peritoneal space when it is, may cause discomfort, delay to treatment, injury to internal organs or vessels.

Distribution

US Nationwide distribution in the states of CA, MN, IN, MI, LA.

States Affected

NATIONWIDE

Quantity Affected

1240

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0887-2026