Class II

Medical Device Recall: Affinity 4 Birthing Bed, Product Codes P3700E100429, P3700E100432, P3700E100433, P3700E100434, P3700E100435, P3700E100436, P3700E100437, P3700E100438, P3700E100439, P3700E100441, P3700E100442, P3700E1

Baxter Healthcare Corporation · May 6, 2025

Reason for Recall

There is a potential for improper crimps on the signal wires of the actuators which could result in a loss of ability to lower the bed or the foot section of the bed.

Product Description

Affinity 4 Birthing Bed, Product Codes P3700E100429, P3700E100432, P3700E100433, P3700E100434, P3700E100435, P3700E100436, P3700E100437, P3700E100438, P3700E100439, P3700E100441, P3700E100442, P3700E100443, P3700E100444, P3700E100445, P3700E100446, P3700E100448, P3700E100449, P3700E100450, P3700E100453, P3700E100455, P3700E100456, P3700E100457, P3700E100459, P3700E100463, P3700E100468, P3700E100472, P3700E100476, and P3700E100477

Distribution

Worldwide distribution - US Nationwide and the countries of BAHRAIN, Canada, Chile, FRANCE, KUWAIT, NETHERLANDS, NORWAY, UNITED ARAB EMIRATES.

States Affected

NATIONWIDE

Quantity Affected

164 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1816-2025

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.