Class II

Medical Device Recall: Baxter PST 500 U, Precision Surgical Table, Part number (product code) 4080300

Baxter Healthcare Corporation · June 28, 2024

Reason for Recall

Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the PST 500 U Precision Surgical Table due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport.

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

States Affected

NATIONWIDE

Quantity Affected

400 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2512-2024

Status: ongoing