Medical Device Recall: Baxter PST 500 U, Precision Surgical Table, Part number (product code) 4080300
Baxter Healthcare Corporation · June 28, 2024
Reason for Recall
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the PST 500 U Precision Surgical Table due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport.
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
States Affected
NATIONWIDE
Quantity Affected
400 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2512-2024
Status: ongoing