Class II

Medical Device Recall: Brand Name: VOLARA System Product Name: VOLARA, HC, NA A Model/Catalog Number: M08594A Software Version: N/A Product Description: VOLARA, HOMECARE, NORTH AMERICA Component: No

Baxter Healthcare Corporation · August 3, 2026

Reason for Recall

Contamination within the release test equipment flow path could have affected the accuracy of the nebulizer flow rate. Affected devices returned to firm and testing confirmed devices meet specifications.

Distribution

US Nationwide distribution in the states of Colorado and Florida.

States Affected

NATIONWIDE

Quantity Affected

2

Source: FDA (openFDA Device Enforcement)

Recall number: Z-3257-2026