Class I

Medical Device Recall: Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product

Baxter Healthcare Corporation · May 30, 2025

Reason for Recall

The Q-link 13 could allow for an improper attachment (false latching) of the Quick-Release Hook used on Sling Bars and other accessories. This could result in a critical injury from a patient fall as the false latched component may initially bear weight but can loosen from the Q-link resulting in a detachment and drop.

Product Description

Hillrom LikoScale component, Product Code 3156232, used with the following adapter kits for patient lifts: 1) LikoScale 200 Accessory - Product Code 3156225; 2) LikoScale 350 Accessory - Product Code 3156228; 3) LikoScale 400 Accessory - Product Code 3156226.

Distribution

Worldwide

Quantity Affected

9839 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2032-2025

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.