Medical Device Recall: Life2000 Ventilator, Product Code: MS-01-0118, which is contained in the Life2000 Ventilator System, with Product Codes BT-20-0002, BT-20-0002A and BT-20-0002AP. The Life2000 ventilation system in
Baxter Healthcare Corporation · May 29, 2024
Reason for Recall
Potential for certain Life2000 ventilator systems to either fail to charge or have intermittent charging behavior due to damage to the battery charger dongle. Damage of the battery charger dongle prevents the ventilator's internal battery from charging.
Product Description
Life2000 Ventilator, Product Code: MS-01-0118, which is contained in the Life2000 Ventilator System, with Product Codes BT-20-0002, BT-20-0002A and BT-20-0002AP. The Life2000 ventilation system includes the Life2000 ventilator which has its individual label.
Distribution
US Nationwide distribution.
States Affected
NATIONWIDE
Quantity Affected
2,510 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2140-2024
Status: ongoing
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.