Class I

Medical Device Recall: Life2000 Ventilator, Product Code: MS-01-0118, which is contained in the Life2000 Ventilator System, with Product Codes BT-20-0002, BT-20-0002A and BT-20-0002AP. The Life2000 ventilation system in

Baxter Healthcare Corporation · May 29, 2024

Reason for Recall

Potential for certain Life2000 ventilator systems to either fail to charge or have intermittent charging behavior due to damage to the battery charger dongle. Damage of the battery charger dongle prevents the ventilator's internal battery from charging.

Product Description

Life2000 Ventilator, Product Code: MS-01-0118, which is contained in the Life2000 Ventilator System, with Product Codes BT-20-0002, BT-20-0002A and BT-20-0002AP. The Life2000 ventilation system includes the Life2000 ventilator which has its individual label.

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

2,510 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2140-2024

Status: ongoing