Class II
Medical Device Recall: Welch Allyn Connex Vital Signs Monitor (CVSM):
Baxter Healthcare Corporation · March 24, 2025
Reason for Recall
There were customer reports of devices which experienced battery-related fires. The investigation found that the devices in question were utilizing third-party, after-market batteries that are not approved by Baxter.
Distribution
Worldwide - US Nationwide distribution.
States Affected
NATIONWIDE
Quantity Affected
228,412 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1558-2025
Status: ongoing