Class I
Medical Device Recall: Welch Allyn Life2000 Ventilation System, REF MS-01-0118, within exterior carton codes: 1) BT-20-0002, 2) BT-20-0002A, 3) BT-20-0002AP, 4) BT-20-0007, 5) BT-20-0011, and 6) RMS010118CP
Baxter Healthcare Corporation · September 12, 2024
Reason for Recall
The Life2000 ventilator may fail to initiate the Low Gas Pressure alarm if the pressure gas source (Life2000 compressor, oxygen cylinder or wall source) is not supplied to the ventilator before initiating therapy.
Distribution
US nationwide distribution.
States Affected
NATIONWIDE
Quantity Affected
10,957 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0046-2025
Status: ongoing