Class II

Medical Device Recall: Access 2 Immunoassay Analyzer, Catalog Number 81600N

Beckman Coulter, Inc. · February 5, 2025

Reason for Recall

Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool used in production (at the supplier), affecting Access 2 sample motors. This could lead to a potential delay in reporting patient test results.

Distribution

Worldwide - US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

873 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1386-2025

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.