Class II

Medical Device Recall: CellMek SPS Sample Preparation System, REF: C44603, with software

Beckman Coulter, Inc. · December 23, 2025

Reason for Recall

Sample preparation system designed to automate staining, lysing, cell washing has a software error that causes samples to be dispensed close to or touching tube bottoms, which could dislodge fluid line, result in sample loss, spillage, make instrument inoperable and will lead to unintentional removal of 60-100 microliters from final sample, may lead to incorrect diagnosis or patient management.

Distribution

Worldwide - US Nationwide distribution in the states of MO, WA, MA, IL, AL, OR, IL and the countries of AB, BE, ES, FR, NL, ON, DE, AT, VIC, AU, GB, SK, MX, CH.

States Affected

NATIONWIDE

Quantity Affected

50

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1235-2026

Status: ongoing