Class II

Medical Device Recall: DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human bo

Beckman Coulter, Inc. · March 5, 2024

Reason for Recall

Beckman Coulter manufacturing discovered cracks around a threaded insert on the bottom of the Internal Wash Buffer Tank that is used within the Wash Buffer Supply Drawer on the DxI 9000.This can result in leaking, leading to delays in reporting patient test results, exposure to physical and chemical hazards, and damage to equipment or the environment.

Product Description

DxI 9000 Access Immunoassay Analyzer, REF C11137, vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.

Distribution

Worldwide - US Nationwide distribution in the state of New Hampshire and the countries of Belgium, Egypt, Italy, Poland, Spain.

States Affected

NATIONWIDE

Quantity Affected

17 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1546-2024

Status: ongoing