Class II

Medical Device Recall: The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Ac

Beckman Coulter, Inc. · September 30, 2024

Reason for Recall

The Access Cortisol assay protocol file (APF) applies an incorrect calibration curve acceptance parameter at the S1 calibrator level that can lead to a high rate of calibration failures which can vary by reagent lot number. Calibration failure would lead to delay of measurement results reporting and potential injuries.

Product Description

The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.

Distribution

Worldwide distribution.

Quantity Affected

149 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0313-2025

Status: ongoing