Class II

Medical Device Recall: CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281

Beckman Coulter Mishima K.K. · January 30, 2026

Reason for Recall

A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment

Distribution

US Nationwide distribution including in the states of NE, WA, PA, CA, AZ, NC, TX, NV, CT, AL, NJ, TN, MT, ID, GA, SC, OH, KS, WI, HI, FL, VA, MI, NH, ME, NY, OR, LA, PR, MN, IL, MD, MO, MS, OK, AR.

States Affected

NATIONWIDE

Quantity Affected

47

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1553-2026

Status: ongoing