Medical Device Recall: DxC 700 AU, REF: B86444, B86446
Beckman Coulter Mishima K.K. · January 30, 2026
Reason for Recall
A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment.
Distribution
US Nationwide distribution including in the states of NE, WA, PA, CA, AZ, NC, TX, NV, CT, AL, NJ, TN, MT, ID, GA, SC, OH, KS, WI, HI, FL, VA, MI, NH, ME, NY, OR, LA, PR, MN, IL, MD, MO, MS, OK, AR.
States Affected
NATIONWIDE
Quantity Affected
142
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1552-2026
Status: ongoing