Class II

Medical Device Recall: BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0116 (30 test kit) BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0104 (6 kit test) The BIOFIRE FILMARRAY Gastrointesti

BioFire Diagnostics, LLC · May 16, 2024

Reason for Recall

Due to increase of false positive Vibrio/Vibrio Cholerae results when using a certain transport medium.

Product Description

BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0116 (30 test kit) BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0104 (6 kit test) The BIOFIRE FILMARRAY Gastrointestinal (GI) Panel is a qualitative multiplexed nucleic acid-based in vitro diagnostic test intended for use with BIOFIRE¿ FILMARRAY¿ Systems. The BIOFIRE GI Panel is capable of the simultaneous detection and identification of nucleic acids from multiple bacteria, viruses, and parasites directly from stool samples in Cary Blair transport media obtained from individuals with signs and/or symptoms of gastrointestinal infection.

Distribution

U.S.: Nationwide - AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VA, Virgin Island, VT, WA, WI, WV, and WY. O.U.S.: Worldwide - AR, AU, AT, BE, BR, CA, CL, HK, CN, CO, CZ, DE, FI, FR, GR, HU, IN, IT, JP, MX, NL, NO, PH, PL, PT, SG, ZA, KR, ES, SE, CH, TW, TH, TR, GB, AL, DZ, AO, AM, BH, BY, BO, BA, BW, BG, CR, HR, CY, CD, DO, AN, EC, SV, EE, FJ, GE, GU, GT, GN, HT, HN, IS, ID, IQ, IE, IL, CI, CI, JO, KE, KW, LV, LB, LT, MO, MY, ML, MR, MD, MN, MA, MM, NP, NZ, NZ, NI, OM, PK, PA, PG, PY, PE, QA, RO, SA, RS, SI, TN, TR, UY, AE, VE, VN, and ZW.

States Affected

NATIONWIDE

Quantity Affected

67,613 kits (or 2,028,309 pouches). Each pouch is used for each individual test.

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2514-2024

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.

BioFire Diagnostics, LLC Medical Device Recall: BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0116 (30 test kit) BIOFIRE FILMARRAY Gastrointestinal (GI) Panel REF RFIT-ASY-0104 (6 kit test) The BIOFIRE FILMARRAY Gastrointesti | SafeCheck