Class I

Medical Device Recall: ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o182, 0173, 0174, 0175,

Boston Scientific Corporation · July 24, 2025

Reason for Recall

Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.

Product Description

ENDOTAK RELIANCE G/SG, transvenous defibrillation lead, implantable cardioverter defibrillator (non-CRT); Model Nos. 0160, 0161, 0162, 0163, 0164, 0165, 0166, 0167, 0170, 0171, o182, 0173, 0174, 0175, 0176, 0177, 0180, 0182, 0183, 0184, 0185, 0186, 0187

Distribution

Worldwide Distribution. US nationwide. International distribution worldwide.

States Affected

NATIONWIDE

Quantity Affected

247,702 US; 344,670 OUS

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2325-2025

Status: ongoing