Class II

Medical Device Recall: EndoVive Safety PEG Kit; Outer Box Material Number (UPN): M00509001; Inner Box Material Number (UPN): M00509000;

Boston Scientific Corporation · February 10, 2026

Reason for Recall

Recall of kits which utilized ChloraPrep Triple Swabsticks from BD. Open or compromised seals on the sterile swabstick packaging, which could result in loss of sterility or drying of the antiseptic solution. This component is placed outside of the sterile barrier of the tray manufactured by Boston Scientific, which contains the majority of the kit components. As such, inclusion of the affected ChloraPrep Triple Swabsticks does not degrade the integrity of the other components within the PEG kit.

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

806 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1551-2026

Status: ongoing