Class I

Medical Device Recall: Expo 5F Selective Angiographic Catheters, MODEL - 5F EXPO AL2 (5PK), REF H74908526982; cardiac catheter

Boston Scientific Corporation · March 21, 2024

Reason for Recall

An increase in complaints related to an inability to advance the guidewire through the lumen of the device for certain batches. There is evidence of the polyurethane layer delaminating and in some instances material detachment in the catheters inner lining.

Distribution

Worldwide distribution. US, US Territories, EMEA, Canada, LATAM and Asia Pacific countries/regions.

Quantity Affected

8715 catheters

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1684-2024

Status: ongoing