Class II

Medical Device Recall: FARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PF21M402, IFU Reference Number 51654495-01

Boston Scientific Corporation · June 11, 2026

Reason for Recall

Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

Distribution

Global Distribution. US Nationwide.

States Affected

NATIONWIDE

Quantity Affected

508,151 units

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2756-2026

Status: ongoing