Class I

Medical Device Recall: RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692, 0693, 0695, 0696

Boston Scientific Corporation · July 24, 2025

Reason for Recall

Affected lead models may experience gradually rising LVSI measurements in association with calcification of the defibrillation lead coils. The issue may lead to early lead replacement. Most serious harm is death or need for cardiac resuscitation.

Distribution

Worldwide Distribution. US nationwide. International distribution worldwide.

States Affected

NATIONWIDE

Quantity Affected

247,702 US; 344,670 OUS

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2327-2025

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.

Boston Scientific Corporation Medical Device Recall: RELIANCE 4-FRONT, endocardial cardioversion/defibrillation and pace/sense lead, Permanent Defibrillator Electrodes; Model Nos. 0654, 0655, 0657, 0658, 0682, 0683, 0685, 0686, 0692, 0693, 0695, 0696 | SafeCheck