Class II

Medical Device Recall: VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201 VXS 63CM/45DEG/D0 SHEATH VXS2112 VERSACROSS SHEATH (63CM 45DEG D0)-J Used for the percutaneous introduction of various

Boston Scientific Corporation · June 10, 2025

Reason for Recall

Potential for hole in the Tyvek layer of the sterile barrier pouch, which may compromise device sterility. Issue could potentially lead to systemic infection and sepsis in worst case scenario. Potential punctures may go undetected in clinical settings due to small size and variable location.

Product Description

VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201 VXS 63CM/45DEG/D0 SHEATH VXS2112 VERSACROSS SHEATH (63CM 45DEG D0)-J Used for the percutaneous introduction of various types of cardiovascular catheters and guidewires to all heart chambers, including the left atrium via transseptal perforation / puncture.

Distribution

Worldwide distribution - US Nationwide and the countries of Canada, Hong Kong, and Japan.

States Affected

NATIONWIDE

Quantity Affected

14,724

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2105-2025

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.

Boston Scientific Corporation Medical Device Recall: VersaCross Transseptal Sheath Kit (VXS) Product Description/IDs VXS0201 VXS 63CM/45DEG/D0 SHEATH VXS2112 VERSACROSS SHEATH (63CM 45DEG D0)-J Used for the percutaneous introduction of various | SafeCheck