Class I

Medical Device Recall: Vivo 45 LS, pressure and volume ventilator capable of delivering continuous or intermittent ventilatory support for patients who require invasive or non-invasive mechanical ventilation, Model/Catalog

Breas Medical, Inc. · July 25, 2024

Reason for Recall

There is a potential for short term (<7 days) elevated levels of formaldehyde emitted from the device into the patient breathing airpath in specific conditions.

Product Description

Vivo 45 LS, pressure and volume ventilator capable of delivering continuous or intermittent ventilatory support for patients who require invasive or non-invasive mechanical ventilation, Model/Catalog Number: 230000 (230016, 230216)

Distribution

US Nationwide.

States Affected

NATIONWIDE

Quantity Affected

8298

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2690-2024

Status: ongoing