Medical Device Recall: MRI system: Vantage Elan, Model: MRT-2020
Canon Medical System, USA, INC. · January 29, 2024
Reason for Recall
For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.
Distribution
Worldwide - US Nationwide distribution including in the states of MD, IL, WV, CA, AZ, PR, CO, GA, PA, WA, UT, OK, KY, IA, IN, TN, CT, WI, OH, AL, MT, NM, TX, MO, MN, OR, ID, WY, NV, AR, FL, SD, MI, NY, NJ, KS, LA, NH, VT, MS, SC, ME and the country of DO.
States Affected
NATIONWIDE
Quantity Affected
1
Source: FDA (openFDA Device Enforcement)
Recall number: Z-1249-2024
Status: ongoing