Class II

Medical Device Recall: MRI systems: Vantage Orian MRT-1550 (MEXL-1550), Vantage Elan MRT-2020 (MEXL-1520), Vantage Titan MRT-1504 (MEXL-1504), Vantage Titan 3T MRT-3010 (MEXL-3010), Vantage Galan 3T MRT-3020 (MEXL-3020)

Canon Medical System, USA, INC. · May 10, 2024

Reason for Recall

The terminal block used to secure the power supply cable for the gradient coil of the MRI system was not secured as designed, which could cause cable disconnection, which could lead to arcing, followed by melting of the cable, ignition, smoke, and fire.

Distribution

US: PA, SC, IA, AL, MI, AR, SD, TX, ID, KY, NY, MN, MD, OH, WI, NV, CT, AZ, IL, OK, CA, FL, GA, OR, IN, NJ, CO, MT, KS, MO, WY, VA, VT, LA, NH, WV, TN, PR

States Affected

AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MD, MI, MN, MO, MT, NH, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, VT, WI, WV, WY

Quantity Affected

153

Source: FDA (openFDA Device Enforcement)

Recall number: Z-2041-2024

Status: ongoing