Class II

Medical Device Recall: Vantage Titan, Model: MRT-1504

Canon Medical System, USA, INC. · January 29, 2024

Reason for Recall

For some MRI systems, it has been found that some of the maximum Spatial Field Gradient (SFG) values displayed in "System Information" of the operation window and some of the maximum SFG values described in the safety manual are lower than the correct values, which may cause a patient with an MR conditional device to feel some discomfort during the MR scan.

Distribution

Worldwide - US Nationwide distribution including in the states of MD, IL, WV, CA, AZ, PR, CO, GA, PA, WA, UT, OK, KY, IA, IN, TN, CT, WI, OH, AL, MT, NM, TX, MO, MN, OR, ID, WY, NV, AR, FL, SD, MI, NY, NJ, KS, LA, NH, VT, MS, SC, ME and the country of DO.

States Affected

NATIONWIDE

Quantity Affected

40

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1250-2024

Status: ongoing