Medical Device Recall: Cardinal Health Monoject, 6ML Enteral Syringe (purple) Low Dose (LDT) - For Use with ENFit Connection System, REF 406SE
Cardinal Health 200, LLC · December 28, 2023
Reason for Recall
The listed lots of Cardinal Health Monoject Enteral Syringes with ENFit (6, 12, 35, 60 mL) have recognition and compatibility issues with certain syringe enteral feeding pumps. As a result, Cardinal Health recommends that they not be used with enteral syringe feeding pumps.
Distribution
Distribution US nationwide and Canada. Foreign distribution updated 9/6/2024 to only include Canada.
States Affected
NATIONWIDE
Quantity Affected
20186 units
Source: FDA (openFDA Device Enforcement)
Recall number: Z-0847-2024
Status: ongoing