Class II

Medical Device Recall: Salem Sump PVC Tubes: Product Code/Product Name: 1180264408 Salem Sump" PVC Gastroduodenal Tube, 10 Fr/Ch (3.3 mm) x 90 cm 1180264416 Salem Sump" PVC Gastroduodenal Tube, 12 Fr/Ch (4.0 mm) x 114

Cardinal Health 200, LLC · September 18, 2025

Reason for Recall

Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.

Product Description

Salem Sump PVC Tubes: Product Code/Product Name: 1180264408 Salem Sump" PVC Gastroduodenal Tube, 10 Fr/Ch (3.3 mm) x 90 cm 1180264416 Salem Sump" PVC Gastroduodenal Tube, 12 Fr/Ch (4.0 mm) x 114 cm 1180264424 Salem Sump" PVC Gastroduodenal Tube, 14 Fr/Ch (4.7 mm) x 114 cm 1180264432 Salem Sump" PVC Gastroduodenal Tube, 16 Fr/Ch (5.3 mm) x 114 cm 1180264440 Salem Sump" PVC Gastroduodenal Tube, 18 Fr/Ch (6.0 mm) x 114 cm 1180264457 Salem Sump" PVC Gastroduodenal Tube, 20 Fr/Ch (6.7 mm) x 114 cm 1180264523 Salem Sump" Polyurethane Gastroduodenal Tube, 14 Fr/Ch (4.7 mm) x 114 cm 1180264531 Salem Sump" Polyurethane Gastroduodenal Tube, 16 Fr/Ch (5.3 mm) x 114 cm 1180264549 Salem Sump" Polyurethane Gastroduodenal Tube, 18 Fr/Ch (6.0 mm) x 114 cm 1180264556 Salem Sump" Polyurethane Gastroduodenal Tube, 20 Fr/Ch (6.7 mm) x 114 cm 1180264895 Salem Sump PVC Gastroduodenal Tube, 20 Fr/Ch (3.3mm), 108 cm 8888264804 Salem Sump PVC Gastroduodenal Tube, 10Fr/Ch (3.3 mm) x 85 cm 8888264820 Salem Sump PVC Gastroduodenal Tube, 12 Fr/Ch (4.0 mm) x 108 cm 8888264846 Salem Sump PVC Gastroduodenal Tube, 14 Fr/Ch (4.7mm), 108 cm 8888264861 Salem Sump PVC Gastroduodenal Tube, 16 Fr/Ch (5.3mm), 108 cm 8888264887 Salem Sump PVC Gastroduodenal Tube, 18 Fr/Ch (6.0mm), 108 cm SKUs not distributed in US

Distribution

Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR, GU and the countries of Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Africa, Kuwait, Luxembourg, Mexico Monaco, Netherlands, New Zealand, Northern Ireland, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Taiwan, UAE, United Kingdom, Vietnam.

States Affected

NATIONWIDE

Quantity Affected

all lots

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0211-2026

Status: ongoing

SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.