Class II

Medical Device Recall: Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Guardian Intraoperative Lead, Medium Model/Catalog Number: REF# 9525 (Medium) Product Description: The Checkpoint Guardi

Checkpoint Surgical Inc · February 24, 2025

Reason for Recall

Adhesive on the back of the lead may not fully encapsulate the metallic portion of the wire and button subassembly, potentially leading to exposed metal. There is a risk of electrical current leakage, which may lead to inconsistent or confusing muscle responses, potentially compromising the effectiveness of the intended use as the surgeon could take an incorrect clinical action.

Product Description

Brand Name: Checkpoint Guardian Intraoperative Lead Product Name: Checkpoint Guardian Intraoperative Lead, Medium Model/Catalog Number: REF# 9525 (Medium) Product Description: The Checkpoint Guardian Intraoperative Leads are single-patient disposable accessories for providing stimulus to a targeted nerve. The accessories are intended for use with any device from the Checkpoint Stimulator/Locator Family. The lead electrode is intended to wrap around nerves that have been surgically exposed, allowing the surgeons to provide hands free stimulus to a targeted nerve.

Distribution

US Nationwide distribution.

States Affected

NATIONWIDE

Quantity Affected

452

Source: FDA (openFDA Device Enforcement)

Recall number: Z-1467-2025

Status: ongoing