Medical Device Recall: Brand Name: Versius Surgical System US Product Name: Vesius Surgeon Console US Model/Catalog Number: V-SS-1020 Software Version: Pratchett 1 version 62.1 Product Description: The Versius Surgical
CMR SURGICAL LIMITED · June 9, 2026
Reason for Recall
Secure Boot was mistakenly not enabled at manufacturing time, which presents a potential cybersecurity risk.
Product Description
Brand Name: Versius Surgical System US Product Name: Vesius Surgeon Console US Model/Catalog Number: V-SS-1020 Software Version: Pratchett 1 version 62.1 Product Description: The Versius Surgical System is a modular, open console, software controlled, electromechanical system designed for surgeons to perform minimally invasive surgery.
Distribution
US Nationwide distribution in the state of California.
States Affected
NATIONWIDE
Quantity Affected
2
Source: FDA (openFDA Device Enforcement)
Recall number: Z-2717-2026
Status: ongoing
SafeCheck provides drug interaction information from FDA-approved product labeling (openFDA). This is for informational purposes only and is not medical advice. Always consult your healthcare provider before making medication decisions.