Class II

Medical Device Recall: MINC+ Benchtop Incubator, Reference Part Number (RPN) K-MINC-2000, Order Number (GPN) G44429

Cook Medical Llc · December 7, 2023

Reason for Recall

The MINC+ device is susceptible to losing temperature control of electrostatic discharge (static electricity) is applied to the lid of the device. This may lead to embryo degeneration necessitating an additional medical procedure for the patient.

Distribution

Domestic distribution to CA IN LA MD NY International distribution to AUSTRALIA AUSTRIA SPAIN FINLAND FRANCE INDIA IRELAND ITALY KUWAIT SWITZERLAND TURKEY UNITED KINGDOM

States Affected

CA, IN, LA, MD, NY

Quantity Affected

14 US; 101 OUS

Source: FDA (openFDA Device Enforcement)

Recall number: Z-0722-2024

Status: ongoing