Medical Device Recall: Newport HT70 Ventilator, REF: HT70M-JP-NA, HT70M-CN-NA, HT70M-ES-EU, HT70M-PT-BR, HT70M-SY-EU, HT70M-WW-EU, HT70M-WW-NA, HT70-SY-EU. Newport HT70Plus Ventilator, REF: HT70PM-ES-EU, HT70PM-ES-NA, HT
Covidien · May 21, 2025
Reason for Recall
Ventilator Printed Circuit Board Assembly may have two separate capacitors that may fail, which may result in the ventilator either shutting down during use, thus necessitating use of an alternate form of ventilation, or the shutdown alert alarm fails to alarm effectively during shut down, which may result in respiratory failure, hypoventilation, low oxygen saturation, hypoxia, treatment delay.
Product Description
Newport HT70 Ventilator, REF: HT70M-JP-NA, HT70M-CN-NA, HT70M-ES-EU, HT70M-PT-BR, HT70M-SY-EU, HT70M-WW-EU, HT70M-WW-NA, HT70-SY-EU. Newport HT70Plus Ventilator, REF: HT70PM-ES-EU, HT70PM-ES-NA, HT70PM-JP-NA, HT70PM-PT-BR, HT70PM-SY-AS, DLHT70PM-WW-NA, HT70PM-SY-NA, HT70PM-SY-UK, HT70PM-WW-EU, HT70PM-WW-UK, HT70PM-SY-EU, HT70PM-WWNA. HT70-2 CONTROL BOARD ROHS X1, REF: GR105796. HT70P CONTROL BOARD ROHS X1, REF: GR105795. HT70 Conversion Kit, REF: GRFSKHT70M-2.
Distribution
Worldwide - US Nationwide distribution including in the states of MA, CT, KS, OH, PR, NH, GA, WI, MS, VA, CO, TX, MN, AL, IA, NY, UT, NM, KY, RI, SD, PA, IA, CA, MO, HI, MI, IL, MT, NJ, TN, SC, FL, IN, AZ, ID, LA, WA, NC, MD, WY, NE, NV, ND, ME, AK, DE, OK, OR and the countries of Japan, Canada, Singapore, Lithuania, Kazakhstan, Slovenia, Colombia, China, Qatar, United Arab Emirates, Greece, Saudi Arabia, Nigeria, Romania, Netherlands, Lebanon, Oman, Jordan, Malawi, India, Croatia, Portugal, Turkey, Brazil, Pakistan, Trinidad and Tobago, Slovakia, Chile, Egypt, Switzerland, South Africa, Thailand, Bangladesh, Myanmar, Peru, Uruguay, Hungary, Argentina, Iran, Islamic Republic Of, Costa Rica, Vietnam, Australia, Guatemala, Mexico, Guyana, Jamaica, Dominican Republic, Bahamas, Bosnia and Herzegovina, South Korea, Tunisia, Ukraine, Uzbekistan, Indonesia, Burkina Faso, Italy, Panama, Sri Lanka, Malaysia, Namibia, Swaziland, Spain, Philippines, Bhutan, Mauritius, Poland, France, Cambodia, Czechia, Israel, Norway, Russian Federation, United Kingdom, Canary Islands, Kenya, Czech Republic, Croatia , Estonia, Maldova, Palestine.
States Affected
NATIONWIDE
Quantity Affected
4842
Source: FDA (openFDA Device Enforcement)